Journal of Electrocardiology
Volume 36, Supplement 1 , Pages 127-132, December 2003

The use of calculated epicardial potentials improves significantly the sensitivity of a diagnostic algorithm in the detection of acute myocardial infarction

  • César Navarro, PhD

      Affiliations

    • Regional Medical Cardiology Centre, Royal Victoria Hospital, Belfast, Ireland, Northern Ireland
  • ,
  • Colum Owens, MRCP

      Affiliations

    • Regional Medical Cardiology Centre, Royal Victoria Hospital, Belfast, Ireland, Northern Ireland
    • Corresponding Author InformationReprint requests: Colum Owens, MRCP, Regional Medical Cardiology Centre, Royal Victoria Hospital, Grosvenor Rd, Belfast, Ireland, Northern Ireland, BT12 6BA
  • ,
  • John Riddell, MRCP

      Affiliations

    • Regional Medical Cardiology Centre, Royal Victoria Hospital, Belfast, Ireland, Northern Ireland
  • ,
  • Anthony McClelland, MRCP

      Affiliations

    • Regional Medical Cardiology Centre, Royal Victoria Hospital, Belfast, Ireland, Northern Ireland
  • ,
  • John McC Anderson, Dphil

      Affiliations

    • University of Ulster at Jordanstown, Belfast, Ireland, Northern Ireland
  • ,
  • Omar Escalona, PhD

      Affiliations

    • University of Ulster at Jordanstown, Belfast, Ireland, Northern Ireland
  • ,
  • Colin Turner, PhD

      Affiliations

    • University of Ulster at Jordanstown, Belfast, Ireland, Northern Ireland
  • ,
  • Jennifer Adgey, MD (FACC)

      Affiliations

    • Regional Medical Cardiology Centre, Royal Victoria Hospital, Belfast, Ireland, Northern Ireland

Abstract 

Inverse electrocardiography can calculate epicardial potentials (EP) from body surface potentials (BSP) taking into account a thoracic volume conductor model (TVCM). Previous studies have shown that a tailored TVCM is superior to a general TVCM in calculating EP. However, construction of a tailored TVCM for a patient in an acute clinical setting is impractical. In this study we used a general TVCM in our EP calculations to determine whether this improves detection of acute myocardial infarction (AMI) using a diagnostic algorithm. BSP were derived from the 80-lead body surface map (BSM). Consecutive patients (n=379) with ischemic type chest pain were recruited. The BSM and a 12-lead electrocardiogram (ECG) were recorded at initial presentation and creatine kinase (CK) and/or CK-MB were measured initially, 12 and 24 hours postsymptom onset. A physician interpreted the 12-lead electrocardiogram and documented ST elevation if present. AMI was defined by the World Health Organization (WHO) criteria. The diagnostic algorithm result for each patient using BSP and calculated EP were documented. AMI occurred in 171 patients. The diagnostic algorithm using BSP identified 106 of these as ST elevation AMI (STEMI) (sensitivity 62%, specificity 80%). The same algorithm using EP identified 133 as STEMI (sensitivity 78%, specificity 80%). Calculated EP improved the algorithm’s diagnostic sensitivity by a factor of 1.25 (P< .001) with no significant difference in specificity. Calculated EP using a general TVCM significantly improves the sensitivity of a diagnostic algorithm based on BSP in detection of AMI with no significant loss in specificity.

Keywords:  Inverse electrocardiography, thoracic volume conductor model, body surface mapping, acute myocardial infarction

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 This work was supported by research fellowships from the Research and Development of the Northern Ireland Health and Social services Agency and the Frances and Augustus Newman Foundation.

PII: S0022-0736(03)00124-9

doi:10.1016/j.jelectrocard.2003.09.036

Journal of Electrocardiology
Volume 36, Supplement 1 , Pages 127-132, December 2003